Scroll for temperature, pressure, inlet flow, and outlet flow. Each chart runs from the first injection through the end of the run.
| Pressure (psi) | Temperature (°C) | Flow rate (SCCM) | Carbon conversion (%) |
|---|
Latest sample loads with the trend.
—
—
—
—
—
—
—
—
—
—
—
—
—
| Place on the drawing | Live value |
|---|---|
| Nitrogen MFC and hydrogen MFC, sccm | A pill sits on each readout. The log has one series, inlet flow (Flow1, AIs-.Module5 Inp2), and does not say which face is nitrogen and which is hydrogen, so both pills stay blank. |
| PT and PG | reactor pressure (Loop4.Process Value, L4PV) |
| Reactor TC and TIC | Reactor (Loop1.Process Value, L1PV) |
| Reactor heater output | Reactor output (Loop1.Output, L1Out) |
| I/P on the line to the separator | Pressure output (Loop4.Output, L4Out) |
| Stirring speed, cooling output, separator temperature, separator heater | Drawn where the SpecView screen shows them. No channel in the live log yet. |
| Solenoid position, NC and NO | Equipment on the screen. Open and closed are not in the live log. |
The packed-bed drawing is not here yet. This view stays blank until that print is added.
Temperature and inlet flow are at reactor exit, after the 1 L dead volume. Point color is that inlet flow. Each line fits conversion against temperature for one inlet flow, rounded to the nearest 100 SCCM. In each equation, y is carbon conversion (%) and x is temperature at exit (°C). Hiding a row drops that injection from this chart, the carbon balance chart, and these fits.
| Pressure (psi) | Temperature (°C) | Flow rate (SCCM) | Carbon conversion (%) |
|---|
Carbon in is the inlet flow times the feed carbon. Carbon out is the product-gas mass flow times the GC carbon (CO + CO2 + CH4). Both are sccm of carbon. The gap between the lines is what conversion treats as reacted carbon. Shaded bands are where that gap has to be corrected before it is called conversion.
Move across the graph to read an injection.
Volume %. The total is checked against 100%.
Total —
Feed gas is over 100%.
Trend line y = mx + b, with peak area as y and gas % as x. Gas % = (peak area − b) / m.
The two Brooks meters are the dual feed, side by side. One is calibrated on N2 and the other on H2. These K values are fixed.
| Gas | N2 Brooks | H2 Brooks |
|---|---|---|
| Hydrogen | 1.008 | 1.000 |
| Carbon monoxide | 0.995 | 0.987 |
| Carbon dioxide | 0.773 | 0.767 |
| Methane | 0.763 | 0.757 |
| Nitrogen | 1.000 | 0.992 |
Saved in the calibration inputs folder on App info.
One row per GC injection. Product-gas cells are the median within 30 seconds. Reactor temperature, pressure, and inlet flow are the nearest SpecView sample. The 1 L dead volume shifts a second set of those reactor values back by 1 / volumetric flow. Gas % uses the calibration curve on this sheet, and a gas with no curve is written as peak area. Carbon conversion uses those percents, this feed, the inlet at exit, and the product-gas mass flow.
These folders are read on every refresh. Change points the program at a different folder.
Run-variable JSON files are saved and loaded from this folder.
Download table writes this file and also sends it to the browser.
Developer: update this table when the steps from the raw files to the injection table change.
| Step | What the program does |
|---|---|
| 1 | Read every *.LOG file in the SpecView folder. Channel names map to these log header names:
|
| 2 | Read livelog.csv in the SpecView folder. Channel names map to these log header names:
|
| 3 | Read every Cole-Parmer file matching flowlog-*.log. The line that starts with =~ and contains “log ” sets the clock, including the date. Column headers in order: pressure, temperature, volumetric flow, mass flow. |
| 4 | Read every datafile_*.rst in the TotalChrom folder, in file-number order. The injection time is the acquisition time (UTC) shifted 5 hours to show the time in EST. Peak area is the largest value between 100 and 1e9 found for Hydrogen, Carbon Monoxide, Carbon Dioxide, and Methane. |
| 5 | For each injection, collect Cole-Parmer seconds from 30 seconds before to 30 seconds after. With at least 10 samples, pressure, temperature, volumetric flow, and mass flow are the median. Otherwise those four cells stay blank. |
| 6 | For each injection, take the nearest SpecView sample. If it is more than 30 seconds away, temperature, pressure, and inlet flow stay blank. Inlet flow is also blank when it is negative or 9000 or higher. |
| 7 | The dead volume from the end of the reactor to the GC detector is 1 L. The gas passes the back pressure regulator and the Cole-Parmer meter on the way. Transit time (min) = 1 / product-gas volumetric flow (LPM). Gas left reactor is the injection time minus that transit time, rounded to the nearest second. Reactor temperature, pressure, and inlet flow at exit are the nearest SpecView sample to that time, with the same blank rules as step 6. A volumetric flow that is blank or 0 leaves these cells blank. The raw columns stay at the injection time. |
| 8 | Apply the curve from Calibration / inputs. Gas % = (peak area − intercept) / slope. These are the adjusted gas percents. They are not scaled to 100%. A gas with no slope is written as its peak area, and that column is labeled area instead of mol%. A missing peak stays blank and counts as 0 in the carbon sum. |
| 9 | Carbon out (mol%) is CO + CO2 + CH4. Feed carbon is the CO + CO2 + CH4 typed on Calibration / inputs, divided by 100. Inlet flow is the inlet at exit, in SLPM, and inlet K is 1. Outlet flow is the product-gas mass flow at the injection, in SLPM. Carbon conversion (%) = 100 × (inlet flow × feed carbon − outlet flow × carbon out / 100) / (inlet flow × feed carbon). No leak is subtracted. A blank inlet at exit, a blank mass flow, or a feed with no carbon leaves the conversion blank. |
| 10 | Write one CSV row per injection to the injection table file. Download table sends that same file. Analysis draws each conversion against reactor temperature at exit. Color is the inlet flow at exit. The table under that graph lists temperature, pressure, inlet flow, and carbon conversion for each of those injections. |
Developed by Dr. Julia Hasty for the Institute of Gas Innovation and Technology at the Advanced Energy Research and Technology Center. 2026